B-54 DELTA/DELTA XL/KAPPA VF8
B T
ECHNICAL DATA
MultiGas Via Scio Module
Alarm limit range: 18 to100 %
Stability of accuracy: ±3 % over an 8-hour interval
Protection against electric
shock:
Type CF
Standards: IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable sections
only, Oxygen analyzers for monitoring patient breathing mixtures); EN
ISO 21647:2009
Note: RTP = Room temperature / pressure 23 °C±3 and ambient barometric pressure
Parameter display: etCO2*, iCO2*, RRc*, et O2, iO2, O2, N2O, etHAL, iHAL,
etISO, iISO, etENF, iENF, etSEV, iSEV, etDES, iDES
NOTE: Asterisks distinguish Scio module parameter labels
from those of the etCO
2
module.
Agents measured: Halothane, Isoflurane, Enflurane, Sevoflurane,
Desflurane
NOTE: With Scio Four plus and Scio Four Oxi Plus the
anesthetic agent identification threshold is max. 0.3 Vol%.
The agent identification threshold of a second agent is 0.4
Vol% or, if Desflurane is the primary agent, one tenth (10%)
of the Desflurane concentration.
Display values: Agents: CO2, O2:Inspired and expired
concentrations (%), plus trend, and waveform
N2O: Inspired and expired
concentrations (%)
Measurement methods: CO2, Agents, N20:Infrared
O2: Paramagnetic sensor
Display range: CO2: 0 to 10 % (resolution 0.1 %)
O2: 0 to 100 % (resolution 1.0 %)
N2O: 0 to 100 % (resolution 1.0 %)
Agents:
Halothane:0 to 10 % (resolution 0.1 %)
Isoflurane:0 to 10 % (resolution 0.1 %)
Enflurane:0 to 10 % (resolution 0.1 %)
Sevoflurane:0 to 11% (resolution 0.1 %)
Desflurane:0 to 24 % (resolution 0.1 %)
Alarm limit resolution: CO2: 0.1 %
O2, N2O: 1 %
Agents: 0.1 %
Full Accuracy after warm-up period complete
Agents:
1, 2, 3
(when RR 60 with I:E of 1:2)
Halothane (up to 8.5 %vol):±(0.15 vol % + 15 % rel.)
Isoflurane (up to 8.5 %vol):±(0.15 vol % + 15 % rel.)
Enflurane (up to 10 %vol):±(0.15 vol % + 15 % rel.)
Sevoflurane (up to 10 %vol):±(0.15 vol % + 15 % rel.)
Desflurane (up to 20 %vol):±(0.15 vol % + 15 % rel.)