Option Description
F2 Set Number of
Points
Enter the number of points. The number of measurement point
directions and number of points determine the total number of locations
to collect data. For example, if the measurement direction is XYZ and
1,024 points are defined, a total of 3,072 locations (1,024 X, 1,024 Y and
1,024 Z) are available in the job. The default is 20.
F3 Set Starting
Point
Enter the first measurement point to display when you collect data. If a
job has 20 measurement points, the starting point could be from 1
through 9,979. If the starting point is 9,979, the second point number
would be 9,980. The default is 1.
F4 Set Coordinate
Sys
Use the up and down arrow keys to select Rectangular (X, Y, Z), Cylindrical
(Radial, Theta, Axial), or Spherical (Radial, Theta, Phi). The default is (X, Y, Z).
Note
ME'scope VES
â„¢
uses X, Y, Z.
F5 Set Meas
Direction
Use the up and down arrow keys to select the direction. The available
directions depend on the coordinate system defined for the job in F2 Set
Coordinate Sys. The default is XYZ. For example, if you selected rectangular
(XYZ), the possible measurement directions are X, Y, Z, XY, XZ, YZ, and
XYZ.
F10 Set Meas
Input
Set the input for the displayed measurement direction (orientation). This
option does not appear if you have only two inputs. If you disable the
input for this direction, "???" displays.
F11 Set Meas
Input
Set the input for the displayed measurement direction (orientation). This
option does not appear if you have only two inputs. If you disable the
input for this direction, "???" displays.
F12 Set Meas
Input
Set the input for the displayed measurement direction (orientation). This
option does not appear if you have only two inputs. If you disable the
input for this direction, "???" displays.
5. Press Enter.
The Edit Job Setup screen appears.
9.6.2 Set the job mode to ODS or Modal
You can only change the mode when the job has no stored data.
Procedure
1.
Activate a job.
2. From the ODS/Modal main menu, press F8 Job Manager > F4 Edit This Job > F7 Set Mode.
Each time you press F7 Set Mode, the mode changes.
3. Press Enter.
ODS/Modal
288 MHM-97432 Rev 7