3-7.1.4 INT DIST/EXT DIST Meter
The internal and external distortion meter are
selectable via softkeys located within the "Me-
ter:" field in the Display zone. Display consists
of a digital readout and bar graph. Distortion is
used to measure the audio quality of the trans-
mitter and receiver modulation. The distortion
meter is selectable via softkey between internal
(coupled from the monitor demodulated signal)
and external (through the DIST input on the front
panel).
The distortion meter operates only at the fixed
frequency of 1 kHz.
3-7.1.5 SINAD Meter
The SINAD m
eter is selected within the "Meter:"
field in the Display zone. Display consists of a
digital readout and bar graph. SINAD is used in
making receiver sensitivity measurements per
EIA specifications using a fixed 1 kHz modula-
tion frequency.
Internal coupling for SINAD is not provided;
input is always via the external BNC port on the
front panel.
NOTE
Optional "C" message or CCITT filters, along
with a 600 ohm load, are available for
selection at the ACVM, SINAD, and Dis-
tortion meter inputs. If your unit is equipped
with one of these, they are selectable thru the
Special Function screen (refer to section 3-
8.5). If one of these is selected an appropriate
message will appear on the message line just
above the softkey labels. CAUTION:
Selection of either filters or load can affect
readings within these meter functions.
3-7.1.6 Counter and Decoding Functions
The following are all accessed via softkey
through the "Meter:" field within the Display
zone. Their inputs are all normally internally
coupled to the monitor demodulated signal for
either direct or "off-the-air" testing. If use of
these functions is needed for an externally
applied signal, the Special Functions screen,
under SYSTEM FUNCTIONS, provides a means
of switching the input of the Counter/decoder
from Internal to External.
These screens contain a "Sensitivity:" field where
MIN or MAX may be selected via softkey. This
provides a means to desensitize the coun-
ter/decoder circuits, if needed to properly mea-
sure very high level signals. Under normal
operation, this field should be set to MAX.
Cursor fields are provided to access the units
low-pass and high-pass baseband filter sections.
Filters may be used to remove unwanted voice
modulation, etc. which may interfere with de-
coding the PL signals. Baseband filters apply
only to internal coupling.
CAUTION
Entries into the high and low pass areas of
this screen write information into the Special
Functions screen and memory. To avoid
problems with other modulation
measurements, make sure settings are set to
original values before leaving these screen
areas.
The display exhibits a digital frequency and
equivalent PL code if applicable. Refer to
Appendix B for valid codes.
PL/PER Counter
This softkey provides a convenient means of
measuring the frequency of Motorola
Private-Line (PL) or any other low frequency
audio tones with 3 digit resolution. Period
measurement makes it possible to measure low
frequencies down to high resolution without the
need for the long gate times associated with
frequency counting.
43

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the General Dynamics R2600 Series and is the answer not in the manual?

General Dynamics R2600 Series Specifications

General IconGeneral
BrandGeneral Dynamics
ModelR2600 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals