Using Instrument Functions
Using limit Testing
To
Create
a
Flat
Limit
Line
In this example, you will create a minimum limit line from 155 MHz to
195 MHz at a level of -3
dJ3.
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
12.
To
access the limit line menu press
(m)
Limi%
Menu
.
‘lb
create a new minimum limit line press Add
limit
Add
Mia
fiae
.
Press
Begin Frequency
, and enter
[ISSJ
MHZ.
Press End Frequency , and enter
(GJ
XFlz
.
Press
Bq@n
Limit , and press
(--3)
&t&r.
Press End Limit , and press
[--3)
En%sr
.
Notice that the analyzer has generated a limit line at about the center of
the display.
To get a less distracting view of the limit line, press
@iKiV]
Maye
Mspl,ay
G~&ti~ule
ON
off
to toggle the graticule off.
‘lb see if your filter meets the minimum limit you have just set up, press
Wior
Nonu
Limit Menu Limit
Tset
On.
The display will now indicate on the display whether the DUT has passed
or failed when compared to the current limits. If you are using the hlter
that was shipped with your instrument, you should see a PASS indicator.
Tb
edit a current limit line and to see what a fail indicator looks like,
press Edit Limit and set the Begin
and End Limit
to 0
dE%
Notice the display now shows the FAIL
indicator.
4-33