I
-
I
-
Making Measurements
Making a Power Measurement using Broadband Detection
Enter
the
Measurement
Parameters
Press the following keys on the analyzer:
c-1
(jMEAs-
POW%@
NOTE
This example measurement uses the default instrument parameters for a power measurement. If your
particular power measurement requires specific parameters (such as frequency range, source power
level, number of data points, and sweep time1 enter them now
3-31
CAUTION
Damage to your analyzer will occur if the receiver input power exceeds
+23
dBm
or 25 Vdc. The analyzer’s source cannot
signilicantly
exceed this
level, however if your DUT has gain, then attenuation on the RF
IN
port
may be necessary. See “When to Use Attenuation and
Amplilication
in a
Measurement Setup” earlier in this chapter for more information.