15-24 SCPI Reference Tables Model 2700 Multimeter/Switch System User’s Manual
Table 15-8
TRACe command summary
Command Description
Default
parame
-
ter* Ref SCPI
TRACe|:DATA Use TRACe or DATA as root command. Sec 6
:CLEar Path to clear the buffer.
[:IMMediate] Clear the buffer.
:AUTO <b> Enable or disable buffer auto-clear. ON
:AUTO? Query state of buffer auto-clear.
:FREE? Query bytes available and bytes in use.
:POINts <NRf> Specify size of buffer (2 to 55000).
100
:POINts? Query buffer size.
:NOTify <NRf> Specify number of stored readings that will set
Trace Notify bit (B6) of measurement event
register (2 to 109999). Must be less than
TRACe:POINts value.
50
:NOTify? Query trace notify value.
:NEXT? Query buffer location for next stored reading.
:TSTamp Path to set timestamp format:
:FORMat <name> Select timestamp format (ABSolute or DELTa). ABS
:FORMat? Query timestamp format.
:TYPE? Query timestamp type for readings presently in
buffer.
:FEED <name> Select source of readings (SENSe[1],
CALCulate[1], or NONE).
CALC
:CONTrol <name> Select buffer control mode (NEVer, NEXT, or
ALWays).
NEV
:CONTrol? Query buffer control mode.
:FEED? Query source of readings for buffer.
:DATA? Read all readings in the buffer.
:DATA:SELected?
<start>, <count>
Specify readings to be returned; specify starting
point (first reading is #0) and number of readings
(count).
*SYSTem:PRESet and *RST have no effect on commands in this subsystem. The listed defaults are defaults set at the factory.

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD