Model 2700 Multimeter/Switch System User’s Manual KE2700 Instrument Driver Examples H-15
Name
Manual
Reference Brief Description
Advance6 None Use Case 6 — Scan 160 channels using 7703 module (see
NOTE):
Type K thermocouple (TC) temperature measurements.
Reference junction – Simulated.
Measurement speed (rate) – 0.01 plc.
Filter – Disabled (no filtering).
Buffer – Store 160 reading strings. Buffer elements include
reading and channel #.
Triggering – Bus control source.
Data retrieval – Continuously store data into buffer. Retrieve
data for every 32 readings.
NOTE: When using a module that has a built-in cold junction,
use the Internal reference junction. Keep in mind that the
buffer and data retrieval will have to be modified to
accommodate the number of scanned channels. Modules
that have cold junction include:
7700 and 7706 modules – 20 available TC channels
7708 module – 40 available TC channels
Advance7 None Use Case 7 — Ten 40-channel scans using 7702 module:
Channel 1 uses an external reference junction.
Measurement speed (rate) – 1 plc.
Filter – Repeat, 25 readings.
Channels 2 through 40 are connected to type K thermocouples.
Measurement speed (rate) – 1 plc.
Filter – Disabled (no filtering).
Open thermocouple detection – Enabled.
Buffer – Store 400 reading strings. Buffer elements include
reading, real time clock, and channel #.
Triggering – Bus control source.
Data retrieval – SRQ when buffer G, H, I, and full.
Table H-2 (continued)
LabVIEW examples

Table of Contents

Other manuals for Keithley 2700

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2700 and is the answer not in the manual?

Keithley 2700 Specifications

General IconGeneral
TypeDigital Multimeter
Measurement FunctionsDC Voltage, AC Voltage, DC Current, AC Current, Resistance, Frequency, Temperature
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Frequency Range3 Hz to 500 kHz
Temperature Range-200°C to +1372°C
DisplayVacuum Fluorescent Display
ConnectivityIEEE-488, RS-232
Channels10
Power Supply100V to 240V AC, 45Hz to 66Hz
DC Current Range10 µA to 3 A
AC Current Range10 µA to 3 A
Temperature MeasurementThermocouple, RTD