EasyManua.ls Logo

Keysight Technologies X Series - Page 253

Keysight Technologies X Series
671 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight N9010A EXA Service Guide 253
Front End Control Troubleshooting
A15 Front End Control Assembly Troubleshooting
Figure 6-7 SW4 Connector Location (Option 532 and 544)
Figure 6-8 Connector Close-up
To verify the control logic, press Mode Preset on the analyzer. Press FREQ,
5 GHz, SPAN, 1 MHz, AMPTD, More 1 of 2, uW Path Ctrl. Standard Path is
selected by default. Carefully place the scope probe tip on the conductive
portion where the green wire goes into the connector, see Figure 6-9.

Table of Contents

Other manuals for Keysight Technologies X Series

Related product manuals