Keysight N9010A EXA Service Guide 97
Instrument Messages
-221 Settings Conflict Errors
-221 Settings conflict; FFT sweep
type is not available while
in Gated Video
The FFT sweep type is not available if you have selected the gated
video function.
-221 Settings conflict; Fixed
marker adjust not available
while Marker Function is on
If a Marker Function is on for a Fixed marker, the marker's reported
value is derived from the function. Therefore, you cannot directly
set the X or Y value of a Fixed marker that has a marker function
turned on.
-221 Settings conflict; Fixed
Marker Y value is not
adjustable with Normalize On
If Normalize is on the Amplitude scale is in dB units, so adjusting
the Y value of a Fixed marker is not possible.
-221 Settings conflict; Freq >
3.6 GHz unavailable while
electronic attenuator
enabled
The electronic attenuator does not function above 3.6 GHz. So if
you have that attenuator enabled, you cannot change the center
frequency so that frequencies above 3.6 GHz are
displayed/measured.
-221 Settings conflict; Frequency
Offset is not available with
Segmented Sweep
The frequency offset feature cannot be used when you are using
the segmented sweep capability.
-221 Settings conflict; Frequency
Offset not available when
Frequency Scale is Log
The frequency offset feature cannot be used when you have
selected a log scale for the frequency axis.
-221 Settings conflict; Function
unavailable with MW Presel
off
You cannot center or adjust the preselector because the
Microwave Preselector is currently off
-221 Settings conflict; Gate
control is Edge for Gated
FFT
You cannot use level triggering to control the gate if you are using
the FFT gating method.
-221 Settings conflict; Gate
control must be Edge for
this Gate Source
You cannot use level triggering to control the gate when you are
using the currently selected gate source.
-221 Settings conflict; Gate is
not available when Marker
Count on
The gate function cannot be used while you have marker count
turned on.
-221 Settings conflict; Gate
Length is not settable in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
-221 Settings conflict; Gate
Length is not settable in
FFT sweeps
The sweep time for FFT sweeps is set by the calculations. So
sweep time settings cannot be adjusted.
-221 Settings conflict; Gate
Method is not compatible
with current Sweep Type
setting
If the Gate is On and you have the FFT Sweep Type manually
selected, then the Gate Method cannot be selected.

Table of Contents

Other manuals for Keysight Technologies X Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies X Series and is the answer not in the manual?

Keysight Technologies X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelX Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals