6 System Functions
System
Notes The values displayed on the screen are only updated upon entry to the screen and not updated while
the screen is being displayed.
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:SYSTem:PON:TIME?
Example :SYST:PON:TIME?
Notes Value is the time since the most recent start-up in seconds.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:CURRent?
Example :CAL:TEMP:CURR?
Notes Value is in degrees Centigrade.
Value is invalid if using default alignment data (Align Now, All required)
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TIME:LALL?
Example :CAL:TIME:LALL?
Notes Value is the elapsed time, in seconds, since the last successful Align Now, All or Align Now, All but
RF was executed.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:LALL?
Example :CAL:TEMP:LALL?
304 EMI Receiver Mode Reference

Table of Contents

Other manuals for Keysight EXA N9010A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight EXA N9010A and is the answer not in the manual?

Keysight EXA N9010A Specifications

General IconGeneral
BrandKeysight
ModelEXA N9010A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals