6 System Functions
System
Notes Value is in degrees Centigrade at which the last successful Align Now, All or Align Now, All but RF
was executed.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TIME:LRF?
Example :CAL:TIME:LRF?
Notes Value is the elapsed time, in seconds, since the last successful Align Now, RF was executed, either
individually or as a component of Align Now, All.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:LRF?
Example :CAL:TEMP:LRF?
Notes Value is in degrees Centigrade at which the last successful Align Now, RF was executed, either
individually or as a component of Align Now, All.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TIME:LPReselector?
Example :CAL:TIME:LPR?
Notes Value is the date and time the last successful Characterize Preselector was executed. The date is
separated from the time by a space character. Returns “” if no Characterize Preselector has ever
been performed on the instrument.
Dependencies In models that do not include preselectors, this command is not enabled and any attempt to set or
query will yield an error.
State Saved No
Initial S/W Revision Prior to A.02.00
EMI Receiver Mode Reference 305