6 System Functions
System
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:LPReselector?
Example :CAL:TEMP:LPR?
Notes Value is in degrees Centigrade at which the last successful Characterize Preselector was executed.
Dependencies In models that do not include preselectors, this command is not enabled and any attempt to set or
query will yield an error.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:AUTO:TIME:OFF?
Example :CAL:AUTO:TIME:OFF?
Notes Value is the elapsed time, in seconds, since Auto Align has been set to Off or Off with Alert. The value
is 0 if Auto Align is ALL or NORF.
State Saved No
Initial S/W Revision Prior to A.02.00
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TIME:RFPSelector:LCONducted?
Example :CAL:TIME:RFPS:LCON?
Notes Values are the date and time the last successful Align Now, 20 Hz – 30 MHz was executed. The date
is separated from the time by a semi-colon character.
State Saved No
Key Path
Visual annotation in the Show Alignment Statistics screen
Mode All
Remote Command
:CALibration:TEMPerature:RFPSelector:LCONducted?
Example :CAL:TEMP:RFPS:LCON?
Notes Value is in degrees Centigrade at which the last successful Align Now, 20 Hz – 30 MHz was
executed.
State Saved No