8-12
8.4.5 Zero the Sensor
Zeroing the sensor aims to eliminate the effect of baseline drift on the readings during the
measurement so as to ensure measurement accuracy.
For microstream CO2 module, a zero calibration is carried out automatically when necessary.
You can also start a manual zero calibration when deemed necessary. To manually start a zero
calibration, enter the [Gas Module Setup >>] menu and then select [Zero]. You do not need
to disconnect the sensor from the breathing system when performing the zeroing.
8.4.6 Calibrate the Sensor
For microstream CO2 module, a calibration should be performed once a year or when the
measured value has a great deviation.
8.4.7 Oridion Information
This trademark is registered in Israel, Japan, German and America already.
Oridion Patents
This device and the CO2 sampling consumables designed for use herewith are covered by
one or more of the following USA patents: 4,755,675; 5,300,859; 5,657,750; 5,857,461 and
international equivalents. USA and international patents are pending.
No Implied License
Possession or purchase of this device does not convey any express or implied license to use
the device with unauthorized CO2 sampling consumables, which would, alone, or in
combination with this device, fall within the scope of one or more of the patents relating to
this device and/or CO2 sampling consumable.