Mindray WATO EX-65 - PEEP Setting Range; Ventilator Performance; Ventilator Monitored Parameters; PEEP Monitored Parameter

Mindray WATO EX-65
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
B-8
Rate 4 to 100 BPM 1 BPM VCV, PCV, PSV
: 4:1 to 1:8 0.5 VCV, PCV, PSV
TIP:TI OFF, 5 to 60% 5% VCV
Finsp 20 to 85 L/min 1 L/min PSVSIMV-VC
SIMV-PC
Trigger Window 5 to 90% 5% SIMV-PC, SIMV-VC
SIMV Rate 4 to 60 BPM 1 BPM SIMV-VC,SIMV-PC
Tinsp 0.4 to 5 s 0.1 s SIMV-VC, SIMV-PC
Inspiratory Trigger
Level
Pressure:
PEEP-20 cmH
2
O to
PEEP-1 cmH
2
O
Flow:
0.5 to 15 L/min
Pressure:
-1 cmH
2
O
Flow:
0.5 L/min
PSVSIMV-VC
SIMV-PC
PSV Insp Termination
Level
5 to 60% 5% PSV, SIMV-VC,
SIMV-PC
Backup Mode Active 5 to 30 s 5 s PSV
PEEP setting range
Type Integrated electronic PEEP
Range OFF, 4 to 30 cmH
2
O; increment:1 cmH
2
O
Ventilator performance
Drive pressure 280 to 600 kPa
Peak flow 100 L/min
Range of flow valve 1 to 100 L/min
Ventilator monitored parameters
MV 0 to 100 L/min
TV 0 to 2500 mL
O2 concentration 18 to 100%
Paw -20 to 120 cmH
2
O
Pmean -20 to 120 cmH
2
O
Pplat -20 to 120 cmH
2
O
: 4:1 to 1:10
PEEP monitored parameter
Range 0 to 70 cmH
2
O