4-15
NOTE
z When SIMV mode, either SIMV-VC or SIMV-PC, is selected, pressure support
ventilation (PSV) mode is used for triggering outside of the trigger window.
Therefore, you also need to set the parameters in PSV mode appropriately,
[Psupp], [Finsp] and [PSV Insp Termination Level].
4.4.5.5 Set Parameters in SIMV Mode
Similar to setting the parameters in VCV and PCV modes, you can use the shortcut keys and
control knob to set the parameters in SIMV mode. The following takes setting of TV as an
example.
1. Select the [TV] shortcut key.
2. Push the control knob and turn it to set [TV] to the appropriate value.
3. Push the control knob to confirm the setting.
4. Set other parameters in this mode in the similar way.
NOTE
z If the parameter value is adjusted outside of the range, the system prompt message
area displays [Parameter Settings Outside the Safety Range].
z Confirm the adjustment of one parameter before adjusting another parameter. If
you want to restore the value before adjustment, you have to reset the parameter
value.
In SIMV (SIMV-VC or SIMV-PC) mode, you also need to set:
[Trigger Window]
1. Select the [Vent Mode] shortcut key [SIMV-VC >>] or [SIMV-PC >>]
[Trigger Window].
2. Push the control knob and turn it to set [Trigger Window] to the appropriate value.
3. Push the control knob to confirm the setting.
4. Select [Ok] to activate the current setting.
5. To cancel the current setting and exit the current menu, select [Cancel],
or push the
Normal Screen key.