9-5
9.5 Make AG Settings
By selecting the [User Setup] shortcut key and then [Gas Module Setup >>], you can make
AG settings described below.
9.5.1 Set Anesthetic Agent
As M-type AG module cannot identify the type of anesthetic agent automatically, you need to
set [Agent] to select the correct type of anesthetic agent before venting the anesthetic agent.
9.5.2 Set Pump Rate
In the [Gas Module Setup] menu, you can select [Pump Rate] and then select
either:[High],[Med] or [Low].
9.5.3 Set O2 Compensation
If the AG module is not integrated with O
2
module, you need to select to set O
2
compensation
based on the actual conditions. Access the [Gas Module Setup >>] menu and select [O2
Comp]. The options include:
[OFF]: when the amount of O
2
in the ventilation gas mixture is less than 30%;
Other options: selects an appropriate value according to the amount of O2 in the
ventilation gas mixture.
If the AG module is integrated with O
2
module, the system calculates compensation directly
using the O2 concentration detected by the O
2
module. In this case, [O2 Comp] is set to
[OFF] permanently and is not user adjustable.
9.5.4 Set Working Mode
The default working mode of the AG module is [Measure] when the anesthesia machine is
turned on. If the current AG module is Standby, you must push the key or select the
[User Setup] shortcut key [Gas Module Setup >>] [Working Mode] [Measure]
to start the AG module. When the anesthesia machine restarts, the AG module automatically
continues with the previously selected working mode.