13.6.3 Install the BIS Module
13.7 Pneumatic Connectors
This anesthesia machine provides two types of connectors —pipeline connectors (for O
2
,
N
2
O and AIR) and cylinder connectors (for O
2
and N
2
O).
For the pipeline connectors, four types of configuration are available:
O
2
O
2
and N
2
O
O
2
and AIR
O
2
, N
2
O and AIR
For the cylinder connectors, three types of configurations are available:
O
2
O
2
and N
2
O
O
2
and O
2
For details, refer to 2.2 Equipment Appearance
WARNING
z Use medical grade gas supplies only. Other types of gas supplies may contain
water, oil, or other contaminants.
z When the central piping system fails, one or more equipment connected may stop
work. Make sure that cylinders are available.
z When gas supplies are cut off, there is still pressure inside the pipeline. Remember
to release the gas inside the pipeline before removing the tube.
z If the [Drive Gas Pressure Low] alarm occurs when the gas supply pressure is
greater than 200 kPa, contact your service personnel or us.