13-6
NOTE
z If it is hard to push the breathing system into or out of the circuit adapter, you
need to apply some lubricant (M6F-020003--- Dupont Krytox high-performance
fluorine lubricating grease)to the seal on the pneumatic connector to reduce
friction.
13.1.4 Install the Bag Arm
1. Align the bag arm with the connector on the breathing system.

Table of Contents

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min