5.1.1 Standby Screen
When the anesthesia machine is not in use for a short period of time, entering standby status
can help save power and extend service life of the machine.
The anesthesia machine enters standby status automatically after start-up. To enter standby
status, you can also push the
key in operating mode and then select [Ok] from the
pop-up menu. The following figure shows the standby screen.
In standby status, the following changes occur to the system:
Displaying monitored parameters and waveforms is disabled. The system is in standby
status.
The ventilator stops supplying gases.
The parameters can be set. When the standby status exits, the system will operate based
on the final settings in standby status.
Physiological alarms are cleared automatically. Technical alarms function normally.
The gas module enters standby status.
To exit standby, push the
key in standby mode and then select [Ok] from the pop-up
menu.