4-8
NOTE
z If the parameter value is adjusted outside of the range, the system prompt message
area displays [Parameter Settings Outside the Safety Range].
z Confirm the adjustment of one parameter before adjusting another parameter. If
you want to restore the value before adjustment, you have to reset the parameter
value.
4.4.3.6 Parameter Range and Default Value in VCV Mode
Parameter Range Step Default
TV 20 to 1500 ml
20 to 100 ml: 5 ml
100 to 300 ml: 10 ml
300 to 1500 ml: 25 ml
500 ml
Rate 4 to 100 BPM 1 BPM 12 BPM
I:E 4:1 to 1:8 0.5 1:2
Plimit 10 to 100 cmH2O 1 cmH2O 30 cmH2O
PEEP OFF, 4 to 30 cmH2O 1 cmH2O OFF
4.4.4 Pressure Control Ventilation (PCV)
4.4.4.1 Description
Pressure control ventilation (hereinafter referred to as PCV) mode is a basic fully-mechanical
ventilation mode. In the PCV mode, each time mechanical ventilation starts, Paw rises
rapidly to the preset Plimit. Then gas flow slows down through the feedback system to keep
Paw constant until expiration starts at the end of inspiration. The tidal volume delivered in
the PCV mode changes based on patient pulmonary compliance and airway resistance.
In the PCV mode, you need to set Plimit to prevent high airway pressure from injuring the
patient.
In the PCV mode, you can also select to set [PEEP] to improve expiration of end-tidal
carbon dioxide and to increase oxygenation of breathing process.