6-2 
 
6.2 Inspect the System 
 
NOTE 
z  Make sure that the breathing system is correctly connected and not damaged. 
z  The top shelf weight limit is 30 kg. 
 
Make sure that: 
1.  The anesthesia machine is undamaged. 
2.  All components are correctly attached. 
3.  The breathing system is correctly connected, and the breathing tubes are undamaged. 
4.  The vaporizers are locked in position and contain sufficient agent. 
5.  The gas supplies are connected and the pressures are correct. 
6.  Cylinder valves are closed on models with cylinder supplies. 
7.  The necessary emergency equipment is available and in good condition. 
8.  Equipment for airway maintenance and tracheal intubation is available and in good 
condition. 
9.  Applicable anesthetic and emergency drugs are available. 
10.  The casters are not damaged or loose and the brake (s) is set and prevents movement. 
11. Make sure the breathing system is locked (in the 
 position). 
12.  The AC mains indicator and the battery indicator come on when the power cord is 
connected to the AC power source. If the indicators are not on, the system does not have 
electrical power. 
13.  The anesthesia machine is switched on or off normally. 
 
6.3 Power Failure Alarm Test 
1.  Set the system switch to the   position. 
2.  Disconnect the AC mains. 
3.  Make sure that the AC mains indicator is extinguished and the battery indicator is 
flashing. Meanwhile, the prompt message [Battery in Use] is displayed. 
4.  Reconnect the AC mains.