3-2
4. MV&TVe alarm key
In case of manual ventilation mode: Push the key to switch off MV and TVe
overrange alarms and apnea alarm. Push the key again to switch on MV and TVe
overrange alarms and apnea alarm.
In case of mechanical ventilation mode: Push the key to switch off MV and TVe
overrange alarms. Push the key again to switch on MV and TVe overrange alarms.
5. Normal screen key
Push the key to close all menus displayed.
6. Standby key
Push the key to enter or exit standby mode.
7. Alarm silence key
To set alarm silence state, push this key to enter 120 s alarm silenced status. The
alarm silence symbol
and 120 s countdown time appear in the upper right
corner of the screen.
To clear alarm silence, push this key again.
8. Operating state LED
On: when the anesthesia machine is operating.
Off: when the anesthesia machine is turned off.
9. AC power LED
On: when the anesthesia machine is connected to the AC power source.
Off: when the anesthesia machine is not connected to the AC power source.
10. Battery LED
On: when the anesthesia machine is equipped with batteries and is connected to the
AC power source, and the batteries are being charged.
Off: when the anesthesia machine is not equipped with batteries or is switched off.
Flash: when the anesthesia machine is being battery powered.
11. Ventilator parameter setup shortcut key(s)
Push the parameter setup shortcut key to change the corresponding setting. Turn the
control knob to change the specific setting and push the control knob to activate the
selected setting.
12. Display screen
Refer to 3.2Display Screen for details.