13-1
13 Installations and Connections
WARNING
z Continuous use of desiccated sodalime may endanger patient safety. Adequate
precautions should be taken to ensure that the sodalime in the sodalime canister
does not become desiccated. Turn off all gases when finished using the system.
z When electrosurgical equipment is used, keep the electrosurgical leads away from
the breathing system, the O
2
sensor and other parts of the anesthesia machine.
Keep backup manual ventilation and simple respirator with mask available in case
the electrosurgical equipment prevents safe use of the ventilator. In addition, make
sure of the correct operations of all life support and monitoring equipment.
z Do not use antistatic or conductive masks or breathing tubes. They can cause burns
if they are used near high frequency electrosurgical equipment.
z This equipment must be installed by the factory authorized engineer.
z This anesthesia machine has waste gas exhaust ports. The operator of the machine
should pay attention to the disposal of the residual breathing gas scavenged.
CAUTION
z The operational environment and the power source of the equipment shall comply
with the requirements in B.2 Environmental Specifications and B.3 Power
Requirements.
13.1 Install the Breathing System
NOTE
z Pay attention to the disposal of the breathing system after equipment use, the
detection of the sodalime in the canister and the anesthetic agent in the vaporizer to
ensure the normal operation of the equipment.