4-16
[Trigger Level]
1. In the SIMV-VC mode, select the [Vent Mode] shortcut key → [SIMV –VC >>] →
[Trigger Level]. Or, in the SIMV-PC mode, select the [Vent Mode] shortcut key →
[SIMV-PC >>] → [Trigger Level]. Or, in the PSV mode, select the [Vent Mode]
shortcut key → [PSV >>] → [Trigger Level].
2. Select [Pressure] or [Flow] for trigger type.
3. Turn the control knob to set [Trigger Level] to the appropriate value.
4. Push the control knob to confirm the setting.
5. Select [Ok] to activate the current setting.
6. To cancel the current setting and exit the current menu, select [Cancel],
or push the
Normal Screen key.
[PSV Insp Termination Level]
1. In the SIMV-VC mode, select the [Vent Mode] shortcut key → [SIMV –VC >>] →
[PSV Insp Termination Level]. Or, in the SIMV-PC mode, select the [Vent Mode]
shortcut key → [SIMV-PC >>] → [PSV Insp Termination Level]. Or, in the PSV
mode, select the [Vent Mode] shortcut key → [PSV >>] → [PSV Insp Termination
Level].
2. Push the control knob and turn it to set [PSV Insp Termination Level] to the
appropriate value.
3. Push the control knob to confirm the setting.
4. Select [Ok] to activate the current setting.
5. To cancel the current setting and exit the current menu, select [Cancel],
or push the
Normal Screen key.