13-25
13.4.2 Fill the Vaporizer
WARNING
z Make sure that the correct anesthetic agent is used. The vaporizer is designed with
the specific anesthetic agent named on it and further indicated by color coded
labelling. The concentration of the anesthetic agent actually output will vary if the
vaporizer is filled with the wrong agent.
13.4.2.1 Pour Fill System
1. Check that the vaporizer concentration control A is in the 0 (zero) position. Check that
the drain screw C is fully tightened.
2. Unscrew the filler cap B.
3. Allow the liquid to flow into the vaporizer slowly. Pay attention to the liquid level
during filling. Stop filling when the maximum level mark is reached.
4. Tighten filler cap B properly.