3-6
3.3.3 Set System Time
1. Select the [Maintenance] shortcut key [User Maintenance >>] [Set System
Time >>].
2. Set [Date] and [Time].
3. Select [Date Format] and toggle between [YYYY-MM-DD], [MM-DD-YYYY] and
[DD-MM-YYYY].
4. Select [Time Format] and toggle between [24 h] and [12 h].
CAUTION
z Changing date and time will affect the storage of trends and log information. It
may also cause loss of data.
3.3.4 Set Language
1. Select the [Maintenance] shortcut key and select [User Maintenance >>].
2. Select [Language] and select the desired language.
3. Restart the anesthesia machine to activate the language setting.
3.3.5 Set Unit
1. Select the [Maintenance] shortcut key and select [User Maintenance >>].
2. Select [Paw Unit] and toggle between cmH
2
O, hPa and mbar.
If the anesthesia machine is configured with CO
2
or AG module, you can set the display units
of FiCO2 and EtCO2. For details, refer to 8 CO2 Monitoring.
3.3.6 Restore Default Configurations
3.3.6.1 Restore the Factory Default Configuration of the Ventilator
To restore the factory default configuration of the ventilator, do as follows:
1. Select the [Maintenance] shortcut key [User Maintenance >>] [Ventilator
Defaults].