WARNING
z Take care to set [CPB] to [ON] because some physiological alarms are not
triggered under this setting. These disabled physiological alarms include: apnea
alarm, Volume Apnea>2 min, Paw too low, TVe too high, TVe too low, MV too
high, MV too low, Rate too high, Rate too low, EtCO2 too low, FiCO2 too low,
EtN2O too low, FiN2O too low, EtHal too low, FiHal too low, EtEnf too low, FiEnf
too low, EtIso too low, FiIso too low, EtSev too low, FiSev too low, EtDes too low
and FiDes too low.
11.7 Set MV&TVe Alarm
1. Push the MV&TVe alarm key when MV&TVe alarm is turned on. The message
[MV&TVe Alarm Off] is prompted and the icon
is displayed on the
screen.
2. Push the MV&TVe alarm key again and the message [MV&TVe Alarm On] is
prompted.
WARNING
z MV&TVe alarm is not triggered when MV&TVe alarm is turned off. Exert care
when using MV&TVe alarm.
11.8 Set Apnea Alarm
In non-mechanical ventilation mode:
1. Push the MV&TVe alarm key when apnea alarm is turned on. The message [Apnea
Alarm Off] is prompted and the icon
is displayed on the screen.
2. Push the MV&TVe alarm key again and the message [Apnea Alarm On] is prompted.
When apnea alarm is turned off, if the anesthesia machine detects breathing waveforms, the
system automatically turns on apnea alarm.

Table of Contents

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min

Related product manuals