4-17 
 
4.4.5.6 Parameter Range and Default Value in SIMV Mode 
 
Parameter  Range  Step  Default  SIMV mode
TV  TV  20 to 1500 ml
20 to 100 ml: 5 ml 
100 to 300 ml: 10 ml 
300 to 1500 ml: 25 ml 
SIMV-VC 
Pinsp  5 to 60 cmH2O    1 cmH2O  15 cmH2O  SIMV-PC 
SIMV Rate  4 to 60 BPM  1 BPM  10 BPM 
Tinsp  0 to 4.5 s  0.1  1.5 s 
Finsp  20 to 85 L/min  1 L/min  60 L/min 
Plimit  10 to 100 cmH2O 1 cmH2O  30 cmH2O 
Psupp  5 to 60 cmH2O    1 cmH2O  15 cmH2O 
PEEP 
OFF, 4 to 30 
cmH2O 
1 cmH2O  OFF 
Trigger Window  5 to 90 %  5 %  25 % 
Pressure  -20 to -1 cmH2O  1 cmH2O  -2 cmH2O 
Trigger 
Level 
Flow  0.5 to 15 L/min  0.5 L/min  3 L/min 
PSV Insp 
Termination Level 
5 to 60 %  5 %  25 % 
SIMV-VC 
SIMV-PC 
 
4.4.6 Pressure Support Ventilation (PSV) 
 
4.4.6.1 Description 
Pressure support ventilation (hereinafter referred to as PCV) mode is an auxiliary breathing 
mode which needs patient’s spontaneous breathing to trigger mechanical ventilation. When 
the patient’s spontaneous inspiration reaches the preset Trigger Level, the ventilator begins to 
deliver gas at the preset Finsp to make Paw rise to the preset Psupp rapidly. After that, the 
ventilator slows down the flow through the feedback system to keep Paw constant. When the 
inspiration flow drops to the preset PSV Insp Termination Level, the ventilator stops 
delivering gas and the patient is allowed to expire, waiting for next inspiration trigger. If 
inspiration is not triggered within the set time (Backup Mode Active), the system 
automatically switches to the backup ventilation mode—PCV. 
 
In the PSV mode, you do not need to set TV. TV depends on the patient’s inspiratory force 
and pressure support level, compliance and resistance of the patient and of the whole system. 
The PSV mode is used only when the patient is driven by reliable breathing because 
breathing must be fully triggered by the patient during ventilation.