B-7
Expiratory resistance
Inspiratory resistance
B.7 Ventilator Specifications
Ventilator parameter setting range
Parameter Setting range Step Operating mode
Plimit 10 to 100 cmH
2
O 1 cmH
2
O All modes
Pinsp PEEP+5 to 60 cmH
2
O 1 cmH2O PCV, PSV,
SIMV-PC
Psupp 5 to 60 cmH
2
O 1 cmH
2
O PSV, SIMV-VC,
SIMV-PC
PEEP OFF, 4 to 30 cmH
2
O 1 cmH
2
O All modes
TV 20 to 1500 ml 20 to 100 ml: 5 ml
100 to 300 ml: 10 ml
300 to 1500 ml: 25 ml
VCV
SIMV-VC
Manual
Mechanical
Pressure drop (kPa)
Flow (L/min)
Manual
Mechanical
Flow (L/min)
Pressure drop (kPa)