B-21
B.14 BIS Module Specifications
BIS Module
Type Single-slot module
Standard IEC60601-2-26
Measurement method Bispectral index, power spectrum analysis
Measured parameters
EEG
BIS: 0 to 100
Calculated parameters
SQI
EMG
SR
SEF
TP
Impedance range 0 to 999 kΩ
Sweep speed 6.25, 12.5, 25 or 50 mm/s
Input impedance >50 MΩ
Noise (RTI) <0.3 µV (0.25 to 50 Hz)
Input signal range ±1 mV
EEG bandwidth 0.25 to 110 Hz
Patient leakage current <10 µA
BIS alarm limits Range Step Unit
BIS High Limit
(low limit + 2) to 100
BIS Low Limit 0 to (high limit – 2)
1 %

Table of Contents

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min