EasyManuals Logo

Mindray WATO EX-65 Operator's Manual

Mindray WATO EX-65
276 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #242 background imageLoading...
Page #242 background image
B-20
Effect of interfering gas on AG measured value
Quantitive effect(%
ABS
)
2)
Gas
Concentration
(%)
CO2 N2O Agent O2
CO2 / / 0.1 0.1 0.2
N2O / 0.1 / 0.1 0.2
AG1) / 0.1 0.1 0.13) 1
Nitrogen ≤78% 0 0 0 0
Xenon <100% 0.1 0 0 0.5
Helium <50% 0.1 0 0 0.5
Ethanol <0.1% 0 0 0 0.5
Acetone <1% 0.1 0.1 0 0.5
Methane <1% 0.1 0.1 0 0.5
Methoxyflurane / Unspecified Unspecified Unspecified Unspecified
1) Multiple agent interference on CO2, N2O and O2 is typically the same as single agent
interference.
2) Maximum quantitive effect of each gas at concentrations within specified accuracy ranges for
each gas. The total effect of all interferences shall not exceed 5%REL of gas concentration.
3) Applicable to AION 03 AG module only, equivalent to the interference of secondary AG to
primary AG.

Table of Contents

Other manuals for Mindray WATO EX-65

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray WATO EX-65 and is the answer not in the manual?

Mindray WATO EX-65 Specifications

General IconGeneral
TypeAnesthesia Machine
ModelWATO EX-65
BrandMindray
Tidal Volume Range20 to 1500 mL
O2 Concentration21% to 100%
I:E Ratio4:1 to 1:8
Ventilation ModesVCV, PCV, SIMV, Manual, Spontaneous
Flow MeterO2, N2O
VaporizerUp to 2 vaporizers
PEEP0 to 20 cmH2O
Gas SupplyO2, N2O, Air
MonitoringAirway Pressure, Flow, Volume
O2 Flush35 to 75 L/min
Respiratory Rate4 to 100 breaths per minute
Power Supply100 to 240 VAC, 50/60 Hz
Fresh Gas Flow0 to 15 L/min