4-4
4.4 Set Ventilation Mode
4.4.1 Set Manual Ventilation Mode
1. Turn the APL valve control to adjust the pressure in the breathing system within the
appropriate range.
2. Set the bag/mechanical ventilation switch to the
position. The ventilation mode
prompt area displays the icon for manual ventilation mode. Besides, the system prompt
message area displays [Manual Vent.].
3. Push the O
2
flush button to inflate the bag if necessary.
In the manual ventilation mode, you can use the APL valve to adjust the breathing system
pressure limit and gas volume in the manual bag. When the pressure in the breathing system
reaches the pressure limit set for the APL valve, the valve opens to release excess gas.
The following figures show the Paw waveform and flow waveform in the manual ventilation
mode.