3-7
2. Select [Ok] from the pop-up menu.
After [Ok] is selected, the following settings restore their default values:
îš„ User screen
îš„ Ventilator parameters
îš„ Alarm limits of ventilator-related parameters
îš„ O2 monitoring source
îš„ Alarm sound volume and key sound volume
îš„ Screen brightness
îš„ Paw display unit
3.3.6.2 Restore the Factory Default Configuration of the Gas Module
If the anesthesia machine is configured with CO
2
or AG module, you can directly restore the
factory default configuration of the corresponding module. For details, refer to 8 CO2
Monitoring and 9 AG and O2 Concentration Monitoring.
3.3.6.3 Restore the Factory Default Configuration of the BIS Module
If the anesthesia machine is configured with BIS module, you can directly restore the factory
default configuration of the corresponding module. For details, refer to 10 BIS Monitoring.
3.3.7 Set the IP Address of Anesthesia Information System
(CIS)
To set the IP address of anesthesia information system (CIS), do as follows:
1. Select the [Maintenance] shortcut key → [User Maintenance >>] → [Set IP
Address >>].
2. In the [Set IP Address] menu, set the correct IP address of the CIS.
3. Select [Ok] to activate the IP address setting.