13-34
13.7.2 Install the Gas Cylinder
For details, refer to 13.5Install/Replace the Gas Cylinder
13.8 CIS Connector
The anesthesia machine can be connected to an anesthesia information system (CIS), which
is to be installed, serviced and updated by Mindray authorized or approved personnel. For
details, refer to the Instructions for Use accompanying the CIS.
13.9 Scavenging
The scavenging assembly is located on the left side of the work table. There are two outlets
labeled AGSS and PEEP as shown below.
1. The PEEP outlet gives off the exhaust gas indoors directly.
2. The outside diameter of the AGSS connector is 30 mm. with 1:20 taper ratio. Please
connect to the AGSS or waste gas disposal system.
WARNING
z The PEEP outlet gives off a small amount of O2 continuously. Do not occlude this
outlet. Otherwise, the anesthetic ventilator cannot work normally.
z Before performing an operation on the patient, equip the anesthesia machine with
anesthesia gas scavenging system which complies with ISO 8835-3 to purify the air
in the operating room.
z If your anesthesia machine is not configured with active AGSS, do not connect the
waste gas exhaust port of the anesthesia machine to the active hospital’s waste gas
disposal system.