9-8
9.9 Scavenge the Sample Gas
To scavenge the sample gas to the waste gas disposal system, depress the metal chip and then
plug the exhaust tube to the ports marked
(sample gas return to the AGSS) on the
anesthesia machine as shown in the above picture.
WARNING
z When using the AG module to perform AG
measurements on the patients who are
receiving or have recently received anesthetic agents, connect the outlet to the
waste gas disposal system to prevent the medical staff from breathing in the
anesthetic agents.
9.10 Calibrate the AG Module
Calibrate the AG module once a year or when the measured value has a great deviation.
Contact us for calibration service.
Metal chip
Exhaust tube