Loading...
14-17
14.3.1 Breathing system
Refer to methods recommended in the table of 14.3 Clean&Disinfect and Re-install the
Breathing System to clean and disinfect the breathing system.
2. Make sure that the breathing system is fully dry before installing it with reference to
13.1.3Install the Breathing system in the reverse order.
14.3.2 Water Collection Cup
1. Refer to the methods recommended in the table of 14.3 Clean&Disinfect and Re-install
the Breathing System to clean and disinfect the water collection cup.
2. Make sure that the water collection cup is fully dry before installing it with reference to
14.2.11 Water Collection Cup in the reverse order:
Align the water collection cup with the matching threaded hole on the breathing system.
Turn the water collection cup counterclockwise to tighten it.
14.3.3 Manual Bag
1. Refer to the methods recommended in the table of 14.3 Clean&Disinfect and Re-install
the Breathing System to clean and disinfect the manual bag.
2. When the manual bag is fully dry, refer to 13.3 Install the Manual Bag to install it.