WARNING 
z  The end of inspiration/expiration connectors which connects the breathing tube 
shall be kept downward to prevent condensed water from entering the breathing 
system. 
 
14.3.10 O2 Sensor 
 
WARNING 
z  Do not put both of the breathing system and the O
2
 sensor in liquid or autoclave 
them. 
z  Water vapor may condense on the surface of the O
2
 sensor, which can result in 
invalid O
2
 concentration measurement. In this case, you need to take out the O
2
 
sensor, remove the water condensed on its surface, and reinstall it into the 
breathing system. 
 
1.  Refer to the methods recommended in the table of 14.3 Clean&Disinfect and Re-install 
the Breathing System to clean and disinfect the O2 sensor. 
2. When the O
2
 sensor is fully dry, refer to 14.2.1 O2 Sensor to install it in the reverse 
order. 
 
14.3.11 AGSS Transfer and Receiving System 
 
14.3.11.1 Filter 
 
Shake the removed filter to get rid of the dust and other contamination until satisfactory 
cleaning effect is achieved. 
 
14.3.11.2 Float 
 
Refer to the methods recommended in the table of 14.3 Clean&Disinfect and Re-install the 
Breathing System to clean and disinfect the float. 
 
14.3.11.3 Spoiler 
 
Refer to the methods recommended in the table of 14.3 Clean&Disinfect and Re-install the 
Breathing System to clean and disinfect the spoiler.