F-2
F.1.3 Sidestream CO2 Module
CO2 Setup Factory default settings
Paw Unit mmHg
Working Mode Measure
Pump Rate High
N2O Comp (%) 0
O2 Comp (%) 0
Des Comp (%) 0
Humidity Comp Wet
F.2 AG Module
AG Setup Factory default settings
Agent AA
Pump Rate Low
O2 Comp OFF
Working Mode Measure
Unit mmHg
Gas Module Alarm Limits
Alarm Level Med
EtCO2 High Limit (mmHg) 50
EtCO2 Low Limit (mmHg) 15
FiCO2 High Limit (mmHg) 4
FiCO2 Low Limit (mmHg) 0
EtN2O High Limit (%) 55
EtN2O Low Limit (%) 0
FiN2O High Limit (%) 53
FiN2O Low Limit (%) 0
EtHal High Limit (%) 3.0
EtHal Low Limit (%) 0.0
FiHal High Limit (%) 2.0
FiHal Low Limit (%) 0.0
EtEnf High Limit (%) 3.0