EasyManuals Logo

Minebea CSD-709 User Manual

Default Icon
218 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
3 Calibration
43
3. Press .
The mV/V value of the currently set zero point appears, and the number in the last digit flashes.
4. Using and , set the mV/V value of the zero point.
CAUTION
When the value set exceeds 2.5 mV/V, flashing appears. When the
value is set is lower than -2.5 mV/V, flashing appears. These outliers are
outside the scope of zero point calibration.
5. Press .
Entry of span
(After zero point entry)
1. Confirm is displayed.
2. Press to display .
CAUTION
Repressing redisplay .
3. Press .
The mV/V value of the currently set span appears and the number in the last digit flashes.
4. Using and , set the mV/V value of the span.
Subtract the input from the measuring unit in the zero calibration from the input from the
measuring unit to which the load equal to the set measuring weight is applied. Set this
calculation result as mV/V value of span.
CAUTION
When the span entered exceeds 3.1 mV/V, flashing appears. When the
span entered is lower than the value used in the zero point adjustment, flashing
appears. Those outliers are outside the scope of span calibration.
5. Press .

Table of Contents

Other manuals for Minebea CSD-709

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Minebea CSD-709 and is the answer not in the manual?

Minebea CSD-709 Specifications

General IconGeneral
BrandMinebea
ModelCSD-709
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals