5 Settings Related to Detection
62
▼ Setting the automatic start level [PF-41]
When you have selected [3] (Automatic start time-specified section) as the desired detecting section,
you should set the threshold automatic start level to P function PF-41. Described below are the
automatic start level setting procedures.
1. Press once to display .
2. Press twice to change the display to .
3. Press .
The P function mode is activated, and the display changes to a two-digit numerical value
prefixed with [PF-].
4. Using and , change the numerical value to and press .
The currently set automatic start level appears.
1 000
For setting procedures with and , see Section 1.6, "Basic key operation".
5. Using and , set the automatic start level.
You can set any value between -99 999 and 99 999, which will be displayed in the load window.
6. Press .
7. Press twice to return to Measurement mode.
5.1.3 Setting the detection operation [PF-01]
Set the detection operation to P function PF-01. Described below are the detection operation setting
procedures.
1. Press once to display .
2. Press twice to change the display to .
3. Press .
The P function mode is activated, and the display changes to a two-digit numerical value
prefixed with [PF-].
Default