EasyManua.ls Logo

Omicron CIBANO 500 PTM - 21.6 Environmental, Mechanical, and Standards Data

Omicron CIBANO 500 PTM
334 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CIBANO 500 PTM User Manual
310 OMICRON
21.6 Environmental conditions
21.7 Mechanical data
Table 21-19: Climate
Characteristic Rating
Temperature
Operating –10 ºC…+55 ºC/+14 ºF…+131 ºF
Storage –30 ºC…+70 ºC/–22 ºF…+158 ºF
Max. altitude
Operating
2000 m/6550 ft, up to 5000 m/16400 ft with limited
specifications
1,2
1. Voltage inputs V IN (CAT III / 300 V) and C (CAT III / 300 V): from 2000 m/6550 ft to 5000 m/16400 ft altitude CAT III
compliance only with half voltage
2. Outputs A (CAT III / 300 V) and B (CAT III / 300 V) and binary inputs A (CAT III / 300 V) and C (CAT III / 300 V): from
2000 m/6550 ft to 5000 m/16400 ft altitude only CAT II compliance or CAT III compliance with half voltage
Storage 12 000 m/40 000 ft
Table 21-20: Mechanical data
Characteristic Rating
Dimensions (w × h× d)
580 mm × 386 mm × 229 mm/22.9 inch × 15.2 inch × 9.0 inch
(w = 464 mm/18.3 inch without handles)
Weight 20 kg/44.1 lb

Table of Contents

Related product manuals