OMICRON 121
Off-service diagnostic methods
17.1.3 Test group execution
Primary Test Manager provides you with a powerful capability for executing test groups. For information
on grouping tests in test groups, see 7.5.2 "Group tests" on page 67. After you set the hardware
configuration of the test set and the test group settings, you can execute all tests of the test group
automatically by clicking the Start all button.
To execute a test group:
1. Group tests in a test group (see 7.5.2 "Group tests" on page 67).
2. Connect the test object to CIBANO 500.
3. In Primary Test Manager, open the test group you want to execute.
4. In the Hardware configuration area, set the hardware configuration. For the hardware configuration
options of CIBANO 500, see Table 17-6: "Hardware configuration options of CIBANO 500" on
page 129.
Figure 17-2: Hardware configuration of the test set for a test group