• Check all cables and connectors between the Omni and the User Switch.
If the trip is still present after the above checks have been made, then the User
Switch may be defective. Contact your service agent.
If it is required to use the Omni without a User Switch being connected, then the
parameter, Switch Detect, should be set to Off. If a User Switch is not used the
responsibility for that decision lies with the healthcare professional.
Because a disconnected User Switch means there is
no emergency stop function, PG Drives Technology
recommend that Switch Detect is this option is
always set to On. PG Drives Technology will accept
no liability for any losses resulting from any other
setting of this parameter.
3.25 Over Pressure
The Omni has detected that the pressure on the Sip and Puff transducer has
exceeded its maximum level. Although a trip screen is not displayed the system
log will show the trip and number of occurrences.
• Check the user is operating the Sip and Puff device correctly.
If the trip is still present after the above check has been made, then the Omni
may be defective. Refer to section 5.
3.26 Gone to Sleep
This occurs when the control system has been left inactive for a time greater
than the parameter Sleep Timer.
An entry is made in the system log each time this occurs.
3.27 Charging
This occurs when the control system detects that a charger is connected to
either Inhibit 1 or Inhibit 3. Refer to section 2.3 for connection details
The Battery charging screen will be displayed during charger connection.
An entry is made in the system log each time this occurs.