If the error is in a non-active module, for example in the ISM but with a drive
Profile selected, then drive will still be possible, however, the diagnostic screen
will appear intermittently.
11.4.9.1 Identified Module
This identifies which module of the control system has registered the problem.
PM Power Module
JSM Joystick Module
ISM Intelligent Seating/lighting Module
OMNI Omni
11.4.9.2 Trip Text
The Trip Text gives a brief description of the trip type.
11.4.9.3 Trip Code
The 4 digit code displayed gives the exact trip that has been recorded.
11.4.9.4 Diagnostic Procedure
Please follow this procedure:
Read and note the Trip Text displayed, the identified Module and
the Trip Code.
Switch off the control system.
Make sure that all connectors on the listed Module and the
wheelchair are mated securely.
Check the condition of the battery.
Find the definition of the Trip Code in the Service Guide, and take
the required action.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the PG Drives Technology R-NET OMNI and is the answer not in the manual?

PG Drives Technology R-NET OMNI Specifications

General IconGeneral
BrandPG Drives Technology
ModelR-NET OMNI
CategoryWheelchair
LanguageEnglish

Related product manuals