produce a ‘band’ of values. During this process it may be beneficial for the user
to look away from the screen. This prevents ‘false’ values where the user may be
trying to attain earlier levels. Once you are satisfied that the user can produce a
soft sip consistently within this band, use the right navigate button on the front of
the Omni to highlight Hard Sip.
The user must now make a series of hard sips to produce a new Hard Sip band.
Ideally there should be as much difference as possible between the Soft Sip
and Hard Sip values. To aid the Omni’s differentiation between these pressures,
the installer should move the Threshold Marker, (shown in the diagram below),
to the middle of the gap between the hardest soft command and the softest
hard command. This is done via the + / - buttons on the front of the Omni.
Once a suitable threshold is set, use the right navigate button on the front of the
Omni to highlight Soft Puff.
Now repeat the sequence of instructions as above for both soft and hard puffs
including the Threshold setting. Once calibration has been performed, use the
right navigate button on the front of the Omni to return to the Calibrate option
in the Sip and Puff branch of OBP.
Please note that all values must be greater than the Deadband setting, for
calibration to be performed successfully. For more details of the parameter
Deadband, refer to the Programming Chapter.
Threshold
ThresholdThreshold
Threshold
Marker
MarkerMarker
Marker