Internal Assemblies — Rear Chassis 4: Repair
105
Replacement:
1 Align the two plastic shields with the screw holes.
2 Install the Power Supply Assembly.
a Position the ground spade connector to your left (see Figure 65,
).
b Secure to the Rear Chassis with the four M3x8 (T10) screws (
).
c Tighten the screws to 6 inch-lb (0.7 N m).
3 Connect Power Supply Assembly (see Figure 65):
a Connect the AC power 2-wire bundle from the Power Supply (
).
b Connect the AC ground spade connector from the Power Supply (
).
c Connect the Power Output Connector (
).
4 Ensure the connectors are fully seated.
5 Install the Fans, see “Fans” on page 102.
6 Secure the Rear Chassis in the normal Position, see “Tilting Rear Chassis on page 91.
7 Close the device. See “Closing the Case” on page 151.
To Complete the Replacement:
Run Performance Verification and Safety testing as described in the “Performance Verification”
chapter.