Operational Check 3: Troubleshooting
35
Operational Check Flow
Follow the screen instructions and respond to the prompts. Once you have answered the last prompt
(Sync button), you can leave the HeartStart XL+ unattended, and the Op Check will complete.
WARNING: Be sure to safely discharge paddles tested during the Op Check, as described in “Operational Check
Te s t s below.
Operational Check Tests
General System Test: Tests the following parameters and features:
internal case temperature
internal voltages of the power supplies and the clock battery
read / write access between PCAs, modules, and programmable components
executable device software and file system integrity, etc.
Therapy Knob: Tests if the Therapy Knob is reading the correct position.