4: Repair Internal Assemblies — Rear Chassis
108
Pivoting Rear Chassis Downward
Pivot the Rear Chassis downward to access assemblies on the top of the Rear and Front Cases.
Preparation
1 Open the Case. See “Opening the Case on page 87.
2 Remove the Rear Chassis Shelf. See “Rear Chassis Shelf on page 93.
3 Position the open HeartStart XL+ display side down, Therapy PCA facing you.
Removal
1 Disconnect cable connectors
, , only from the Therapy PCA, see Figure 69 and Tab le 3 3.
Figure 69 Therapy PCA Connections
Table 33 Therapy PCA – Rear Chassis Connections
Ref. # PCA Mark Description Connects To Disconnect By
J10, Apex narrow spade connector, white wire
Therapy Port
extension
Pull, wiggle, use pliers if
needed
J8, Sternum wide spade connector, red wire
J16, White wide spade connector, white wire
Therapy
Capacitor
Do not disconnect yet
J15, Red narrow spade connector, red wire
J4 ribbon cable Processor PCA Release latches at edges
of connector, pull