4: Repair Internal Assemblies — Rear Chassis
110
Therapy PCA
Removing the Therapy PCA involves disconnecting many cables and removing many screws. Take your
time and be methodical.
CAUTION: Never touch the Therapy PCA surface with your fingers.
Keep the replacement Therapy PCA in the antistatic pouch until ready to install.
Preparation
1 Open the Case. See “Opening the Case on page 87.
2 Remove the Rear Chassis Shelf. See “Rear Chassis Shelf on page 93.
3 Position the open HeartStart XL+ display side down, Therapy PCA facing you.
4 Pivot the Rear Chassis downward. See “Pivoting Rear Chassis Downward” on page 108.
Removal
1 Disconnect all remaining Therapy PCA cables,
, see Tab le 3 4 and Figure 71.
You have already disconnected cables
, , when pivoted the Rear Chassis downward.
a Disconnect cables
, , .
TIP: Use your fingernails to disconnect the battery data cable
. Do not use any tools to avoid
crushing the fragile connector
.
b Remove both Fan wires (
, ) from the Rear Chassis grommet.
c Remove the Therapy PCA AC Power supply wire
from the Rear Chassis opening.
Table 34 Therapy PCA Connections
Ref. # PCA Mark Description Connects To Disconnect By
J10, Apex narrow spade connector, white wire
Therapy Port
extension
Pull, wiggle, use pliers if
needed
J8, Sternum wide spade connector, red wire
J16, White wide spade connector, white wire
Therapy
Capacitor
J15, Red narrow spade connector, red wire
J4 ribbon cable Processor PCA Release latches at the
connector sides, pull
J5 3-wire bundle Battery PCA Gently pull, wiggle
J1 2-wire bundle Power supply
Push on latch to release,
pull, wiggle
J2 4-wire bundle Battery PCA
J12 3-wire bundle
Fans Pull, wiggle
J13 3-wire bundle