5: Performance Verification Performance Verification Procedures
172
Figure 107 illustrates the HeartStart XL+ circuitry. Compare with Class I test diagram shown in
IEC 62353.
Test Cables
Safety Test Cables can provide a convenient way to perform electrical safety tests. You may order these
cables via the Philips SPS organization, see Tabl e 38 :
AC Mains (Ground Leakage)
Leakage through earth (ground) wire of AC power cord.
Normal Condition (Open ground), both AC line connections intact
–Should be <
300 μA (UL, 120 VAC).
–Should be <
500 μA (IEC 240 VAC).
Record as “aaa”.
Single Fault Condition (Open ground, open neutral), one AC line connection open
–Should be <
1000 μA. Record as “bbbb”.
Figure 107 HeartStart XL+ Safety Test Diagram
Table 38 Safety Test Cables
Tool Code # Part Number Description
TC352 451261026041 Cable for SpO
2
safety test
TC356 451261026081 Cable for ECG safety test
TC366 451261026191 Cable for Defib. Pads/Paddles safety test
APMP
L(N)
N(L)
L
N
PE
MD
Legend:
AP: Applied Part
L: Line
MD: Measuring Device
MP: Mains Part
N: Neutral
PE: Protective Earth
NOTES:
Protective earth does not connect to any accessible parts.
Check accessible conductive parts for touch current (enclosure leakage current), but
not for protective earth resistance because they are not protectively earthed.
plastic box